Function:Put the spliced crystal bar and flitch together on YX-6F orientation instrument, let the bottom of flitch rely on base surface of regular instrument, open the optical shutter to do rechecking measurement. YX-6F can do recheck measurement of X axis for spliced crystal bar, the result can be compared with print result by YX-6D, if happen difference, can remove the crystal bar from flitch and splice again, make sure splice accuracy of crystal bar before cutting.
X-ray tube |
copper target with anode grounding, force air cooling |
Tube voltage |
30KVP,switch closed under full voltage |
Tube current |
0---5mA,continuous adjustable. |
Input power |
Not over 0.3KW |
Orientation accuracy |
±30″ |
Worm gear running error |
Not over 30″ |
Counter rotation angle |
-10°------+100° |
Sample rotation angle |
-10°------+50° |
Counter tube |
Geiger tube, voltage of 600-----1100V |
Time constant |
0.1、0.4、3second; 3levels adjustable |
Protection |
High voltage interlock protection, temperature protection |
slit |
4′、5′、6′ |
Outer dimension |
1200mm×650mm×1100mm |
Equipment weight |
250KG |
Sample diameter |
Silicon single crystal Φ3---Φ8inch silicon wafer; |
Measured crystal bar length |
Silicon single crystal Φ3—8inch;500mm length crystal bar |
|